MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/25/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. Udi not available for this system at time of filing. Device manufacturing date is unavailable. A medtronic representative went to the site to test the equipment. The representative replaced the computer, bulkhead connector, and network cable. The navigation system then passed the system checkout and was found to be fully functional. No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used outside of procedure. It was reported that the 3d scans were not transferred.
 
Manufacturer Narrative
Additional information: udi updated, manufacturing date updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7624280
MDR Text Key111838481
Report Number1723170-2018-02905
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 07/12/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-