Model Number 8637-40
Device Problems Volume Accuracy Problem (1675); Insufficient Flow or Under Infusion (2182)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/20/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a device manufacturer representative regarding a patient receiving morphine at an unknown dose and concentration via an implantable infusion pump. The indication for use was spinal pain. It was reported that the actual reservoir volume (arv) was greater than the expected reservoir volume (erv); arv was 17 and erv was 4. 1. The caller stated that the patient has had 2 adjustments since his last refill and has felt better with each one. No symptoms were reported and it was noted that a dye study was planned. No further complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7625893
MDR Text Key111916458
Report Number3004209178-2018-14018
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 06/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/14/2016
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/20/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured05/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-