• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC COLIBRI II HANDPIECE INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problems Device Stops Intermittently (1599); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/07/2018
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the small battery drive device worked intermittently and was making a loud noise. It was not reported if there were any delays to the surgical procedure. A spare device was available to complete the surgery. There was patient involvement reported. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOLIBRI II HANDPIECE
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ 4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ 4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7626165
MDR Text Key112367827
Report Number8030965-2018-54483
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/08/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number532.101
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-