• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW DISC HUM 4X100MM RT FLANGED C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW DISC HUM 4X100MM RT FLANGED C Back to Search Results
Catalog Number 114905
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Date 06/13/2018
Event Type  Injury  
Event Description
Revision surgery - due to the implants being loose. The surgeon put in longer sizes.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to loose implants. The original surgery date was not provided or could be established, therefore the length of in-vivo service could not be determined. There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery. There are no reported pre-existing patient health conditions. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The device was disposed of at the hospital and not made available to djo surgical for examination. This complaint evaluation is limited in scope since the part associated with this investigation was not returned to djo surgical for evaluation. No information submitted with the complaint that would indicate a material, design, or manufacturing issue with the explanted part. A review of the device history record (dhr) was not conducted since the item and or lot number was not provided or determined during the complaint evaluation. No additional information was obtained to assist in the event identification. As of 27-jun-2018, no records have been forwarded by zimmer-biomet concerning this event. Should zimmer-biomet provide additional information concerning this complaint, the complaint will be re-opened and a further review shall be conducted. The root cause of this complaint was a revision surgery due to loose implants. There are multiple factors that may contribute to an event that are outside of the control of djo surgical. Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDISCOVERY ELBOW
Type of DeviceDISC HUM 4X100MM RT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7626617
MDR Text Key111931855
Report Number1644408-2018-00599
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue Number114905
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/27/2018
Is This a Reprocessed and Reused Single-Use Device? No

Patient Treatment Data
Date Received: 06/21/2018 Patient Sequence Number: 1
Treatment
114700; 114813; 114813; 114905; 114813; 114905
-
-