• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8519
Device Problem Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a hole in the tubing of a clearlink system continu-flo solution set below the drip chamber.The hole in the tubing was discovered during an unspecified process step and was unknown if a patient was involved.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed and a hole in the tubing was noted.Pressure and clear passage under water testing was performed and a leak was observed.The reported condition was verified.The cause was incorrect material handling during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7627119
MDR Text Key112075391
Report Number1416980-2018-03774
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048994
UDI-Public(01)00085412048994
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/27/2022
Device Catalogue Number2C8519
Device Lot NumberR17K27085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received07/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-