MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems Calibration Error (1078); Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/24/2017
Event Type  malfunction  
Manufacturer Narrative
Inspected 1 opened/used enlite sensor and perform continuity resistance test.Sensor failed per specifications.
 
Event Description
The customer reported via phone call that they had inaccurate sensor readings that triggered the threshold suspend.The sensor glucose was reading 60, but blood glucose value was 191 mg/dl.The customer tested again and calibrated at 141 mg/dl.The customer stated they received calibration error and change sensor alerts.The customer removed the sensor and indicated it was not bent.The sensor was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7628210
MDR Text Key112200637
Report Number3004209178-2018-84065
Device Sequence Number1
Product Code OZO
UDI-Device Identifier20643169646432
UDI-Public(01)20643169646432(017)20171013
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/13/2017
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG1SSJR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received05/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age54 YR
Patient Weight185
-
-