Model Number S7
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. Unique device identification (udi) is unavailable. A manufacturer representative went to the site to test the equipment. The system performed as intended. The navigation system passed the system checkout and was found to be fully functional. Device manufacturing date is unavailable.
 
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure. It was reported that the system was not holding its charge when unplugged from the wall. There was no patient present during this event.
 
Manufacturer Narrative
The uninterruptible power supply (ups) for the navigation system was returned to the manufacturer for analysis. The ups was found to be fully functional with no problem found. The reported event could not be duplicated by medtronic personnel. The batteries for the ups were returned to the manufacturer for evaluation. Testing found that the unit powered down immediately after being fully charged. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
Additional information: udi and manufacture date received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
jack edell
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7628254
MDR Text Key111986244
Report Number1723170-2018-02938
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 08/15/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-