Model Number MMT-715NAL |
Device Problems
Device Inoperable (1663); Device Displays Incorrect Message (2591); Battery Problem (2885)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
It was reported that their insulin pump received failed battery test alarm.The customer¿s blood glucose level was 103 mg/dl.Customer reports failed battery test alarm was occur again after troubleshooting.The customer was also advised revert to back up plan and treat per healthcare professional¿s instructions.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
Device was received with normal operating currents and passed self-test, unexpected restart error test and displacement test.No unexpected low battery, weak battery, battery out limit and failed battery test alarms noted during testing.
|
|