JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Device Problems Failure to Unfold or Unwrap (1669); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/04/2018
Event Type  malfunction  
Manufacturer Narrative
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that intraocular lens had unfolding issue with patient contact.There was no patient injury, no incision enlargement.Reportedly another lens with same model and same diopter was used as replacement lens for the patient.No further information provided.
 
Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 8/7/2018.Device returned to manufacturer? yes.Device evaluation: the device was received at the manufacturing site for investigation.Visual inspection with unaided eyes showed that the lens case contained no lens.Further examination also showed no lens contained the box.No testing can be performed since the lens is not physically returned.The reported complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There were no non conformances or discrepancy with respect to the manufacturing process.Historical data analysis: the search in complaint history revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7629435
MDR Text Key112552831
Report Number3011852734-2018-00107
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474531758
UDI-Public(01)05050474531758(17)210711
Combination Product (y/n)Y
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 01/01/2005,08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/11/2021
Device Model NumberZCB00
Device Catalogue NumberZCB0000220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received08/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-