Model Number 6070111 |
Device Problems
Insufficient Flow or Under Infusion (2182); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/02/2018 |
Event Type
malfunction
|
Event Description
|
Vial2bag would only transfer 1/2 of the medication in the vial into the bag.Tip of device was well below the level of the drug however only 1/2 of the medication would transfer.The medication was squeezed back into the vial several times with the same result.Another vial2bag was used and worked appropriately.The patient was not negatively impacted.The device was not retained.
|
|
Event Description
|
Vial2bag would only transfer 1/2 of the medication in the vial into the bag.Tip of device was well below the level of the drug however only 1/2 of the medication would transfer.The medication was squeezed back into the vial several times with the same result.Another vial2bag was used and worked appropriately.The patient was not negatively impacted.The device was not retained.
|
|
Event Description
|
Vial2bag would only transfer 1/2 of the medication in the vial into the bag.Tip of device was well below the level of the drug however only 1/2 of the medication would transfer.The medication was squeezed back into the vial several times with the same result.Another vial2bag was used and worked appropriately.The patient was not negatively impacted.The device was not retained.
|
|
Search Alerts/Recalls
|