Model Number MMT-551NAP |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Customer reported via phone call that the insulin pump had unexpected restart.The customer¿s blood glucose level was 81 mg/dl at the time of the incident.Insulin pump received multiple unexpected restart alarm.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Device passed the displacement, unexpected restart error test and self test.No unexpected restart error noted during test.
|
|