• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The navigation system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used for a catheter placement procedure.The issue occurred intraoperatively while navigating and cause no delay to the surgery.It reported that the distance to target did not show in a shunt case.The surgeon tried changing crosshair movement with no resolution to the issue.The distance to target was not shown on the guidance view.The surgery was completed with navigation and there was no impact on patient outcome.
 
Manufacturer Narrative
Additional information: medtronic received information that the reported issue has not recurred for subsequent shunt placement procedures.The logs for the navigation system were reviewed by medtronic personnel.It was noted that the instrument was suspected to not be in the electromagnetic field and that the surgeon was off plan by 6.9 millimeters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7633304
MDR Text Key112174908
Report Number1723170-2018-02951
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight95
-
-