Model Number MMT-551NAL
Device Problems Fluid Leak (1250); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/28/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event. The device has been returned, but not yet evaluated. Further information will follow once the analysis has been completed. No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system. The customer¿s blood glucose level was 86 mg/dl at the time of the incident. The customer also reported that the insulin was squirting during manual prime process. The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan. The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Insulin pump passed the displacement test but compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk. Unable to perform the occlusion, prime and excessive no delivery test due to compromised force sensor system alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name530G INSULIN PUMP MMT-551NAL
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7636637
MDR Text Key112326290
Report Number3004209178-2018-84928
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMMT-551NAL
Device Catalogue NumberMMT-551NAL
Device Lot NumberA4551NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-