Pump passed the sleep current measurement, active current measurement, and displacement test.Pump trace download analysis confirmed the pump alarmed power error 25 and power loss alarm.The power management tool confirmed pump error 25 and power loss alarms were triggered when the loaded voltage was less than 3.5 volts for four consecutive hours due to connector resistance on the electrical board.After disconnecting and reconnecting the internal battery connector on the electrical board, the pump was monitored and functioned properly.
|