Model Number N/A |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported four screwdrivers have broken springs and can no longer turn a screw when it is engaged in bone.When you turn the proximal knob it just spins.This issue did not affect any surgeries, it was identified while stocking trays.There was no patient involvement.
|
|
Event Description
|
This follow-up report is being submitted to relay additional information.
|
|
Manufacturer Narrative
|
Reported event was unable to be confirmed due to limited information received from the customer.Functional testing and inspections could not be performed due to the products not being returned and no photographs being provided.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|