ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
The report of the autopulse platform (sn (b)(4))'s display screen backlight not functioning was confirmed during functional testing of the platform upon initial power up.The root cause was attributed to a defective processor board.As part of routine service during testing, the platform was further examined and a battery clip is missing.This physical damage found during visual inspection is a characteristic of mishandling of the device.This observation is not related to the reported event.After replacement of the battery clip and processor board, the platform was further tested and passed all functional testing criteria and met all required specifications.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) was observed with no lcd backlight on the user control panel.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7639641
MDR Text Key112442184
Report Number3010617000-2018-00668
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000079
UDI-Public00849111000079
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-