INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; DOUBLE FENESTRATED GRASPER Back to Search Results
Model Number 420189-09
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2018
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was missing from clevis.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
 
Event Description
It was reported that during a da vinci-assisted ileal conduit procedure, the double fenestrated grasper instrument had "insufficient" grip.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Type of Device
DOUBLE FENESTRATED GRASPER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale
Manufacturer Contact
tabitha reed
950 kifer rd.
sunnyvale 
4085232100
MDR Report Key7639853
MDR Text Key112847839
Report Number2955842-2018-10302
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874111581
UDI-Public(01)00886874111581(10)N10170929
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number420189-09
Device Lot NumberN10170929 36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Date Manufacturer Received06/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-