• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 98400-000024
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 06/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Physio-control evaluated the customer¿s device and verified the reported issue. Physio replaced the system pcb assembly to resolve this issue. After other unrelated repairs were completed, proper device operation was observed through functional and performance testing. The device was returned to the customer for use. Physio further evaluated the removed system pcb assembly and determined that the pcb was the cause of the reported issue; however further component cause could not be determined.
 
Event Description
The customer contacted physio-control to report that their device displayed the word ¿service¿ across the display approximately 30 seconds after being powered on. This is indicative of a potential device lock-up. There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7641677
MDR Text Key112587886
Report Number0003015876-2018-01024
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 06/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number98400-000024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-