It was reported that during a da vinci-assisted cholecystectomy procedure, a possible "spark" or "flash" was observed near the illuminator cord connection.
The customer stated that a smell of smoke was then noted at the time the incident occurred.
At that time, the surgeon made the decision to complete the procedure using traditional laparoscopic techniques.
There was no report of any patient harm, adverse outcome or injury.
An intuitive surgical, inc.
(isi) field service engineer (fse) was dispatched to the site.
The fse found no evidence of smoke, fire or melted materials associated with the da vinci surgical system.
Review of the site's system log for the reported procedure date found multiple occurrences of system error code # 48244.
The fse concluded that the system error code 48244 was associated with the illuminator.
To resolve the issue, the fse replaced the illuminator.
The illuminator provides the light which is transported to the system endoscope and projected onto the surgical site.
|