• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
The device remains in service at this time.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) and a non-boston scientific right ventricular (rv) lead exhibited impedance measurements of greater than 2000 ohms, noise, and oversensing.The noise was reproducible, but was felt to have been due to myopotential oversensing.No pacing inhibition of greater than two seconds was observed.The device was reprogrammed and the patient would continue to be monitored.No adverse patient effects were reported.The device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7646680
MDR Text Key112715672
Report Number2124215-2018-11415
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/09/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-