Model Number 3662
Device Problem Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 06/13/2018
Event Type  Injury  
Manufacturer Narrative

The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by (b)(4) on 02 june 2017. The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported. The manufacturer defers to the patient's physician regarding medical history.

 
Event Description

It was reported the ipg was unable to connect to the external device following total hip replacement surgery despite placing the ipg into surgery mode. Troubleshooting was unable to resolve the issue. Surgical intervention may be pending to address the issue.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 7 ELITE
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
tanja hoffman
6901 preston road
plano, TX 75024
9723098520
MDR Report Key7647538
MDR Text Key112648517
Report Number1627487-2018-06080
Device Sequence Number1
Product Code LGW
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/26/2018
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received06/28/2018
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date10/20/2018
Device MODEL Number3662
Device LOT Number5678849
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/13/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/20/2016
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C

Patient TREATMENT DATA
Date Received: 06/28/2018 Patient Sequence Number: 1
Treatment
MODEL 3186 (2), SCS LEAD
-
-