Model Number MMT-1715K
Device Problem Inaccurate Delivery (2339)
Patient Problem Hyperglycemia (1905)
Event Date 05/09/2018
Event Type  Injury  
Manufacturer Narrative

Insulin pump passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime or seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test at 0. 08710 inches.

 
Event Description

The customer reported via phone call that the customer was hospitalized due to the high blood glucose level (b)(6) 2018 and the customer was alleging the insulin pump for under delivery of the insulin. The customer¿s blood glucose level at the time of the incident was over 600 mg/dl. The customer¿s current blood glucose level was 240 mg/dl. The customer was treated with insulin. Customer was assisted with the troubleshooting. The reason why customer was alleging the insulin because the basal was not kicking in the insulin pump will be returned for the analysis.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7649741
MDR Text Key112702258
Report Number3004209178-2018-86564
Device Sequence Number1
Product Code OZO
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation PATIENT
Type of Report Initial
Report Date 06/29/2018
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received06/29/2018
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date06/12/2018
Device MODEL NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device LOT NumberHG26RZJ
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer06/19/2018
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/01/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured12/08/2017
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient TREATMENT DATA
Date Received: 06/29/2018 Patient Sequence Number: 1
-
-