Insulin pump passed sleep current measurement, active current measurement, self test and displacement test.
No unexpected low battery alarm noted during testing.
Pump trace download analysis confirmed the pump alarmed power loss alarm, replace battery alert, replace battery now alarm.
The power management tool indicated abnormal behavior of the lithium backup battery loaded voltage (loaded vlith) as shown on the power management graph and confirmed power loss alarm, replace battery alert and replace battery now alarm due to connector resistance j6 /pcb 1.
After disconnecting and reconnecting the internal battery connector on j6/pcb 1, the pump was monitored and functioned properly.
Unit was received with partially detached reservoir retainer, scratched case, minor scratched lcd window and faded serial number label.
|