Model Number 97702
Device Problems Unable to Obtain Readings (1516); Impedance Problem (2950); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/08/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis results were not available at the time of this report.A follow-up report will be sent when analysis is completed.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who had an implantable neurostimulator for spinal pain.It was reported that there were a couple of electrodes that were out of range.When impedance values were ran at 80hz, 210pw, 3.2v and xxx (500), it read 9-10 days recharge interval for calculation.No symptoms or further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Continuation of d11: product id 3777-60 lot# serial# (b)(4) product type lead product id 3777-60 lot# serial# (b)(4) product type lead product id 97715 lot# serial# (b)(4) implanted: (b)(6) 2018 product type implantable neurostimulator supplemental to update h.6 codes, code 11, 3221 no longer applies, the codes 213, 825 and 4310 apply to the implantable neurostimulator (serial number (b)(4) while the codes 122, 452 and 4315 apply to the electrical lead (serial number (b)(4) h3: analysis of the electrical lead (serial number (b)(4) found that the lead body conductor was broken and the anchor site was unknown.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7650080
MDR Text Key112715693
Report Number3004209178-2018-14638
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109513
UDI-Public00643169109513
Combination Product (y/n)N
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2016
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Date Manufacturer Received08/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age67 YR
-
-