(b)(4).Batch # p93g0m.Investigation summary: the device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged, not all functional testing could be performed with the generator.The jaws were found attached to the instrument.In addition no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
|