GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0313
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a capsule which failed to attach.There was no harm to the patient, no intervention was required, and a repeat procedure was performed.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.No lubrication was used to facilitate placement of the capsule and the delivery system will be returned for investigation.
 
Manufacturer Narrative
Additional information: evaluation summary one bravo ph capsule delivery device was received for investigation.The capsule was not returned.Visual inspection did not reveal any damage and appears to have functioned within specification.Functional testing could not be performed because this is a single use device and once the capsule is delivered it cannot be functionally tested.Investigation conclusion for the failure to attach could not be reliably determined.A review of the device history record was performed and indicates that the product was released meeting finished product specifications.We were unable to confirm the customer¿s report.Corrected information:(evaluation code-method, evaluation code-result, evaluation code-conclusion).If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Corrected information:if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
5920 longbow drive
boulder, CO 80301
7632104064
MDR Report Key7654343
MDR Text Key112852771
Report Number9710107-2018-00813
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101361688
UDI-Public7290101361688
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0313
Device Catalogue NumberFGS-0313
Device Lot Number39166Q
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-