• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL VELA; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL VELA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 06/13/2018
Event Type  malfunction  
Manufacturer Narrative
Vyaire file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.The customer reported the suspected main printed circuit board assembly (pcba) is available for analysis and a return good authorization has been issued.At this time, vyaire has not received the suspected main pcba for evaluation.
 
Event Description
The customer reported while using the vela ventilator; the device displays transducer fault alarms.The customer determined the most likely cause of the reported event is the main printed circuit board assembly.The customer reported there is no patient involvement associated with the event.
 
Manufacturer Narrative
Results of investigation: the vyaire failure analysis laboratory received the suspect component for investigation.An investigation was performed and identified the root cause of the reported issue was due to the pt802 to have recorded faults in the events log and to have failed calibration.This issue will be internally investigated within vyaire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 north riverwoods blvd
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
palm springs CA 92262
Manufacturer Contact
mindy faber
26125 north riverwoods blvd
mettawa, IL 60045
8727570116
MDR Report Key7655884
MDR Text Key112996425
Report Number2021710-2018-09034
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVELA
Device Catalogue Number16532-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/27/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-