• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) ILAB ULTRASOUND IMAGING SYSTEM, VERSION 1.3; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) ILAB ULTRASOUND IMAGING SYSTEM, VERSION 1.3; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number H749ILAB100CARTD0
Device Problem Retraction Problem (1536)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr.: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is device not returned as the complaint did not include returned device review and lacked the objective evidence or descriptive conditions of the event required to determine what could have contributed to the event.(b)(4).
 
Event Description
Same case as mdr id: 2134265-2018-06134 and 2134265-2018-06135.It was reported that automatic pullback failure occurred.An ilab ultrasound imaging system was used in conjunction with an unknown imaging catheter and a pullback sled to view the target lesion.After setup of the catheter was performed, it was noted that during first auto pullback the image and pullback stopped.After that, it was unable to perform automatic pullback.Subsequently, the picture became green during manual pullback.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILAB ULTRASOUND IMAGING SYSTEM, VERSION 1.3
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7656010
MDR Text Key113050035
Report Number2134265-2018-05946
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
PMA/PMN Number
K072517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749ILAB100CARTD0
Device Catalogue NumberILAB100CARTD
Device Lot Number0000006800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/09/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-