DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The hose was replaced, and the unit was returned to service.Patient information could not be obtained due to country privacy laws.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported the unit had a large leak in manual ventilation mode.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7656471
MDR Text Key112996652
Report Number2112667-2018-01313
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-