BARD PERIPHERAL VASCULAR, INC. G2 FILTER SYSTEM - FEMORAL; VENA CAVA FILTER Back to Search Results
Catalog Number RF310F
Device Problems Extrusion (2934); Patient Device Interaction Problem (4001)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/26/2018
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: a complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for perforation of the ivc as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported through the litigation process that sometime post filter deployment, filter limb perforation was identified.No reported attempts were made to remove the filter.The current status of the patient is unknown.
 
Manufacturer Narrative
A complete manufacturing review could not be conducted for the investigation as the lot number is unknown.The device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for perforation of the ivc as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.A review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event Description
It was reported through the litigation process that sometime post filter deployment, filter limb perforation was identified.No reported attempts were made to remove the filter.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary:the device was not returned for evaluation.The medical records included images.The image review was documented in the medical records.Medical records were provided and reviewed.Approximately seven years and one month post filter deployment, computed tomography revealed there was an inferior vena cava filter in place.Approximately one year and nine months later, computed tomography revealed an anterior strut extended 7 mm beyond the anterior wall of the inferior vena cava and contacted the posterior wall of the third portion of the duodenum.There was no evidence of bowel perforation.Therefore, the investigation is confirmed for perforation of the inferior vena cava (ivc).Based upon the available information, the definitive root cause is unknown.Labeling review: instructions for use: potential complications: there have been reports of complications, including death, associated with the use of vena cava filters in morbidly obese patients.A review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: b5,d4(expiry date: 04/2012),g3,g4.H11: d1,d4,h6(result and conclusion).H11:section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient in conjunction with or before bariatric procedure.At some time post filter deployment, it was alleged that the filter struts perforated.The device has not been removed and there were no reported attempts made to retrieve the filter.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 FILTER SYSTEM - FEMORAL
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7657023
MDR Text Key113004088
Report Number2020394-2018-01062
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K062887
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRF310F
Device Lot NumberGFTB3632
Was Device Available for Evaluation? No
Date Manufacturer Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age53 YR
-
-