The patient was undergoing a coil embolization procedure in the splenic artery using a lantern delivery microcatheter (lantern).During the procedure, the physician placed the lantern in the target location using a non-penumbra sheath after navigating through tortuous anatomy.However, the physician then was unable to advance a coil through the lantern and, therefore the lantern was removed.Upon removal, the physician noticed the lantern was kinked.The physician then placed a neuron 6f 070 delivery catheter (neuron 070) to address the extreme tortuosity, and the procedure was completed using a new lantern and the same sheath.There was no report of an adverse effect to the patient.
|
Results: the lantern had ovalizations at approximately 145.5 cm, 147.0 cm, 148.0 cm and 149.5 cm from the hub.The device was kinked approximately 149.0 cm from the hub.The lantern could not be flushed with the bleach solution due to a clot of blood.A 0.023¿ mandrel was inserted to loosen the clot but was unable to advance.Therefore, the device could not be functionally tested.Conclusions: evaluation of the returned lantern revealed a kink on the distal end of the catheter.If the device is forcefully manipulated within tortuous anatomy, the device may become kinked.Further evaluation revealed ovalizations on the distal shaft of the catheter.If the device is forcefully gripped during attempts advance it into a parent device, the catheter may become ovalized.These damages may have contributed to the inability to advance the embolization coil during the procedure.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|