Device evaluation: a correlation between the device and the suspected thrombus could not be conclusively determined through this evaluation.The evaluation of returned pump did not reveal any device related issues.The pump was returned assembled with the driveline severed approximately 11 inches from the pump housing and the distal portion was not returned.The sealed inflow conduit (inlet tube, flex section, and inlet elbow) and the sealed outflow cannula were returned attached to their respective pump ports.Upon disassembly, visual inspection of the pump blood-contacting surfaces revealed no evidence of adhered depositions or thrombus formations.Examination of the pump bearings, rotor, and blood-contacting surfaces under a microscope did not reveal any anomalies related to wear or damage.Electrical continuity testing of the returned portion of the driveline did not reveal any discontinuities or shorts.Functional testing of the pump under normal operating conditions revealed normal pump power consumption and pressure values comparable to the data recorded during the manufacturing process.The pump operated as intended.Device thrombosis and hemolysis are listed in the instructions for use as potential adverse events that may be associated with the use of the heartmate ii left ventricular assist system.A review of the device history records revealed the device met applicable specifications.No further information was provided.The manufacturer is closing the file on this event.
|