• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 229047
Device Problems Signal Artifact/Noise (1036); Under-Sensing (1661)
Patient Problem No Patient Involvement (2645)
Event Date 06/07/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was unable to confirm the reported sensing issues.Analysis did note that the nine volt battery returned with the device needed to be replaced.The battery was replaced.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the analyzer had sensing noise on the ventricular channel and no sensing on the atrial channel.The analyzer was returned for service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7658383
MDR Text Key113014704
Report Number2182208-2018-01226
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00721902256011
UDI-Public00721902256011
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number229047
Device Catalogue Number229047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-