• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO SPACEMAKER; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO SPACEMAKER; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number SMBTTRNDX
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during dissection on a laparoscopic inguinal hernia repair procedure, the obturator handle snapped off.They remove the trocar to remove the obturator and then replaced the trocar to resolve the issue in order to complete the case.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the first returned product noted that the inflation syringe was not received.The insufflation bulb not was received.The obturator top was disengaged.The lock collar, trocar balloon and dissector balloon appeared intact.Pmv performed functional testing; the hole was covered, and the dissector balloon was inflated using a test insufflation bulb, no leaks were detected.The trocar balloon was inflated using a test syringe, no leak was detected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the disengaged obturator top may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEMAKER
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key7660323
MDR Text Key113272736
Report Number2647580-2018-03296
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521545304
UDI-Public10884521545304
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151356
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model NumberSMBTTRNDX
Device Catalogue NumberSMBTTRNDX
Device Lot NumberP6M0061X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
-
-