MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Volume Accuracy Problem (1675); Insufficient Flow or Under Infusion (2182); Device Displays Incorrect Message (2591)
|
Patient Problems
Pain (1994); Therapeutic Effects, Unexpected (2099)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8835, serial#: (b)(4), product type: programmer, patient.Other relevant device(s) are: product id: 8835, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was receiving at morphine, unknown concentration at unknown dose via intrathecal drug delivery pump for non-malignant pain.It was reported that bolus was unexpectedly declined.Patient felt like she was not getting relief and she was in terrible pain.Patient reported her patient programmer (ptm) was malfunctioning and not giving her the bolus when it should.The ptm was giving her normal lockouts of 3 hours when it should deliver a bolus.Today she had refill and the healthcare professional (hcp) took out medication that should have been used up and to call the manufacturer for replacement ptm.Patient was having uplink issue and wanted a ptm replacement.Patient had a lot of medicine left in pump due to not being able to get her bolus's per patient.No further complications were reported.
|
|
Manufacturer Narrative
|
(b)(4).Ptm returned for analysis.Analysis found no significant anomaly.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|