Model Number 8637-20
Device Problems Volume Accuracy Problem (1675); Insufficient Flow or Under Infusion (2182); Device Displays Incorrect Message (2591)
Patient Problems Pain (1994); Therapeutic Effects, Unexpected (2099)
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 8835, serial#: (b)(4), product type: programmer, patient.Other relevant device(s) are: product id: 8835, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient who was receiving at morphine, unknown concentration at unknown dose via intrathecal drug delivery pump for non-malignant pain.It was reported that bolus was unexpectedly declined.Patient felt like she was not getting relief and she was in terrible pain.Patient reported her patient programmer (ptm) was malfunctioning and not giving her the bolus when it should.The ptm was giving her normal lockouts of 3 hours when it should deliver a bolus.Today she had refill and the healthcare professional (hcp) took out medication that should have been used up and to call the manufacturer for replacement ptm.Patient was having uplink issue and wanted a ptm replacement.Patient had a lot of medicine left in pump due to not being able to get her bolus's per patient.No further complications were reported.
 
Manufacturer Narrative
(b)(4).Ptm returned for analysis.Analysis found no significant anomaly.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7660488
MDR Text Key113457101
Report Number3004209178-2018-14978
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169100831
UDI-Public00643169100831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2017
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-