Model Number URF-P7
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4). (b)(4) sent this device to a third party laboratory for additional microbiological testing. The additional microbiological testing indicated no microbial growth for the all channels of this device. Therefore, the testing result cleared the french guideline. Also, omsc reviewed the manufacturing history of this device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device. [first time; (b)(6) 2018] instrument channel: unspecified microbes (1cfu / 100ml). [second time; (b)(6) 2018]. Instrument channel: unspecified microbes (3cfu / 100ml). The device was a brand-new device that was not used on patient yet. The device had been manually reprocessed using peracetic acid. There was no report of patient infection associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO FIBERSCOPE
Type of DeviceURETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key7662164
MDR Text Key113628307
Report Number8010047-2018-01277
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation
Type of Report Initial
Report Date 07/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-P7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-