Device is a combination product.(b)(4).Device evaluated by mfr.: synergy ous mr 4.00 x 16mm stent delivery system was returned for analysis.An examination of the crimped stent revealed proximal stent damage.Proximal struts 1- 5 were damage with struts lifted and pulled in a distal direction.The remainder of the stent was undamaged.The undamaged crimped stent outer diameter was within the specification.The balloon cones were reviewed and there were no issues to note.The balloon wings were tightly wrapped and evenly folded.A visual and tactile examination of the hypotube revealed multiple kinks.An examination of the shaft polymer extrusion found no issues.There were no signs of damage or strain to the port bond.The tip was visually examined and damage was noted.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|