• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. RECOVERY FILTER SYSTEM VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. RECOVERY FILTER SYSTEM VENA CAVA FILTER Back to Search Results
Catalog Number RF048F
Device Problems Migration or Expulsion of Device (1395); Malposition of Device (2616); Extrusion (2934); Patient Device Interaction Problem (4001)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Malfunction  
Manufacturer Narrative

Manufacturing review: as the lot number for the device was not provided, a manufacturing review could not be performed. Investigation summary: the device was not returned for evaluation. Images and medical records were not provided for review. Therefore, the investigation is inconclusive for tilt, migration, and perforation of the ivc as no objective evidence has been provided to confirm any alleged deficiency with the filter. Based upon the available information, the definitive root cause is unknown. Labeling review: a review of product labeling documents (e. G. Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.

 
Event Description

It was reported through the litigation process that sometime post filter deployment, the filter allegedly had multiple struts perforating the inferior vena cava with struts contacting the posterior wall of the duodenum. Allegedly the filter also had inferior migration with angular tilt. There were no reported attempts made to retrieve the filter. The current status of the patient is unknown.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameRECOVERY FILTER SYSTEM
Type of DeviceVENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key7665220
MDR Text Key113288186
Report Number2020394-2018-01117
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK031328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation OTHER CAREGIVERS
Type of Report Initial,Followup
Report Date 07/18/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/05/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberRF048F
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/16/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-