The insulin pump passed operating current, displacement test but the compromised force sensor system alarm occurred during the basic occlusion test due to loose/protruded drive support disk.
Unable to perform the occlusion, prime/compromised force sensor system and excessive no delivery test or verify prime/fill anomaly and rewind anomaly due to compromised force sensor system alarm.
The following physical damage was noted: cracked case at the display window corner, minor broken reservoir tube lip, minor scratched lcd window, missing end cap sticker, cracked case reservoir tube window corner and minor broken battery tube threads.
No moisture damage on the lcd board, mother board, interface board, rf board, motor, vibrator motor and battery tube assembly during visual inspection noted.
|