Model Number 2090AA4 |
Device Problems
Computer Software Problem (1112); Device Displays Incorrect Message (2591); Operating System Becomes Nonfunctional (2996)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: analysis was able to confirm the reported programmer crashing and displaying error messages.The hard drive was replaced.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the programmer crashed while attempting to interrogate an implantable pulse generator (ipg) and then displayed an error message.The programmer was rebooted and then showed an additional error message.The programmer was returned for service.There was no patient involvement.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|