• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Signal Artifact/Noise (1036); Failure to Calibrate (2440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) of the intra-aortic balloon pump (iabp) involved in the event was reviewed.There were no non-conformances in the production dhr related to the reported event.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and was unable to verify the problem.The fse replaced the fiber optic module.The iabp unit passed all calibration, functional and safety tests performed.The unit was returned to the customer and cleared for clinical use.
 
Event Description
It was reported that upon insertion of a sensation plus intra-aortic balloon (iab), the cardiosave intra-aortic balloon pump (iabp) displayed an altered waveform and would not calibrate.A false cyclical/ spiked artifact displayed throughout the fiber-optic arterial pressure waveform with no alarm.The balloon was left in and the staff used the fluid lumen as an alternate arterial lumen.There were no effects to the patient.The patient was being transported to another facility; therefore, the staff did not attempt to arrange for return of the iab.No adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7674697
MDR Text Key113814444
Report Number2249723-2018-01140
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-