Model Number S7
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. Testing revealed that the surgeon monitor was replaced to resolve the issue. The system then passed the system checkout and was found to be fully functional. The surgeon monitor was returned to the manufacturer for analysis. Analysis found that the returned monitor has nicks and scratched on the monitor housing. When connected to a known good system the display was blank. Analysis found that the reported event was related to a electrical issue. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Event Description
Medtronic received information regarding a navigation device being used for a cranial resection procedure. The event occurred intraoperatively during the navigate task and delayed surgery by less than one hour. It was reported that the surgeon monitor started showing dark lines across the screen and was getting distorted while navigating. The site unplugged and reseated the monitor cables and the monitor turned off. There was a high pitch noise indicating power to the monitor but no video. The site used an external monitor to finish the case. The surgery was completed with navigation and there was no known impact on patient outcome.
 
Manufacturer Narrative
Additional information: udi and manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7675427
MDR Text Key113623094
Report Number1723170-2018-03280
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-