Catalog Number 9735665
Device Problems Erratic or Intermittent Display (1182); Intermittent Communication Failure (4038)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2018
Event Type  malfunction  
Manufacturer Narrative
A system checkout was not required as resolution of the issue was confirmed at the time of the event.No parts have been returned to the manufacturer for evaluation.Part not returned.
 
Event Description
Medtronic received information regarding a navigation device being used for a sacroiliac and thoracolumbar procedure.The issue occurred intraoperatively during the navigate task and delayed the surgery by less than one hour.It was reported that the camera cart monitor went dark and indicated no connection.The monitor cycled power and returned to normal display after 10 seconds.The surgery was completed with navigation and there was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tiberious mose
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7676085
MDR Text Key113623390
Report Number1723170-2018-03288
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial
Report Date 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9735665
Date Manufacturer Received06/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age76 YR
Patient Weight91
-
-