COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number FS-OMNI-ACRO-35-260
Device Problems Migration or Expulsion of Device (1395); Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/15/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: (2) cook soehendra dilators, unknown models.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The subassembly device history record for the wire guide could not be reviewed because a lot number was not provided.The device history record for the lot number said to be involved could not be reviewed because a lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use instruct the user to do the following: "for best results, wire guide should be kept wet." failure to flush the wire guide can result in damage to the wire guide.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all fusion preloaded with acrobat wire guide sphincterotomes are subjected to a visual inspection to ensure device integrity.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
As reported to customer relations: a stent was in place from a previous endoscopic retrograde cholangiopancreatography (ercp).The fusion pre-loaded with acrobat wire guide sphincterotome was used to access the duct next to the existing stent.The wire guide was left in place in the duct and the sphincterotome was removed.A snare was passed through the endoscope and used to remove the stent.A soehendra dilator was passed over the in place acrobat wire and the common bile duct dilation was successful.The next size up soehendra dilator was passed over the acrobat wire.There was difficulty getting the dilator out of the endoscope channel.It was then noted that the acrobat wire coating had started to peel and was obstructing the dilator from passing over the wire.The dilator could not be advanced.The physician had to pull out both the wire guide and dilator from the duct and endoscope.A new acrobat wire guide had to be opened and used to gain access to the bile duct again [lost wire guide access].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key7685280
MDR Text Key114541289
Report Number1037905-2018-00312
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFS-OMNI-ACRO-35-260
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-