Model Number 382400 |
Device Problems
No Audible Alarm (1019); Appropriate Term/Code Not Available (3191)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The customer did not report any specific complaint.On (b)(6) 2018, the unit was triaged and service found that there was no audio.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo epump was performed for the reported condition of, ¿service found that there was no audio.¿ the unit was triaged and the reported condition was confirmed.When the pump was powered on startup tones were not audible.Internal inspection revealed a damaged component u14 in the buzzer circuit.A trend has been identified and a corrective action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
|
|