Model Number MMT-1715K
Device Problems Crack (1135); Moisture Damage (1405)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that insulin pump had moisture underneath display screen and in battery compartment.Customer reported that insulin pump was cracked.Customer's blood glucose was unknown.Customer was advised that insulin pump would need to be replaced.Insulin pump is expected to be returned for failure analysis.
 
Manufacturer Narrative
Unit passed displacement test and self test.Traces of moisture were found at the electronic and motor assemblies per visual inspection.Unit received with cracked case on the back at the battery tube area and battery tube threads.Unit received with fading serial number label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7686915
MDR Text Key114154022
Report Number3004209178-2018-91394
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840(017)20200828
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/28/2020
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG2172N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age41 YR
Patient Weight155
-
-