• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypoxia (1918); No Code Available (3191)
Event Date 06/18/2018
Event Type  Injury  
Event Description
It was reported that a vns patient recently had her device adjusted.She began experiencing coughing and respiratory issues.The coughing occurs roughly every 3 minutes.She had the parameters reduced back to the previous settings.The coughing and the respiratory issues did not resolve.She taped the magnet over the device and the issue resolved.She stated that her daughter does have underlying respiratory issues.Follow-up from the company representative later provided that the caregiver had placed the magnet over device for over 12 hours, and at first it helped.It was stated her oxygen saturation was going lower, possibly as a result of the coughing.They were concerned the vns was causing respiratory distress.The company representative provided she observed the patient cough when they removed the magnet.Additional relevant information has not been received to-date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7687277
MDR Text Key113959870
Report Number1644487-2018-01182
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/21/2018
Device Model Number106
Device Lot Number204027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received06/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
-
-