Model Number MMT-723NAS
Device Problems Device Displays Incorrect Message (2591); Power Problem (3010); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/06/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was unknown at the time of the incident.The customer also reported that the pump received compromised force sensor system alarm when they changed the reservoir.The customer stated that they had received error alarm and it occurred during the rewind and prime sequence.It was also stated that the pump had a whining noise and then shut off.The customer was advised that the pump needs to be replaced.The customer was advised to discontinue use of the insulin pump and to revert to the backup plan as per health care professional instructions.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7690018
MDR Text Key114224940
Report Number3004209178-2018-92068
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00613994743930
UDI-Public(01)00613994743930
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Device Lot NumberA3723NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Date Manufacturer Received06/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient Weight195
-
-