MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1711K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1711K |
Device Problems
No Display/Image (1183); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/13/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump had a flashing display and was beeping.Blood glucose level at the time of the incident was not provided.Troubleshooting was not completed.The insulin pump was returned for analysis.
|
|
Manufacturer Narrative
|
Insulin pump was received with a constant blank flashing white light on the display and constantly beeping due to vertical cracked lcd controller.Unable to verify missing segments/partial display or perform the functional testings including the self-test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test and displacement test due to insulin pump flashing white light on the display.
|
|