• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751613
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Corneal Clouding/Hazing (1878)
Event Date 06/22/2018
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A doctor reported multiple post operation reaction(haziness) with approximately 20 patients.Sample of the air and fluid results indicate a culture developed.The customer refused to share any further information on affected patients and product.
 
Manufacturer Narrative
The lot complaint history was reviewed.This is the third complaint for the finished goods lot and for this issue for this lot.The device history record shows the product was released per specifications.No sample was received.Therefore, no visual inspection or functional testing could not be conducted.The root cause of the customer's complaint could not be established as a sample has not been received.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.After a thorough investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned and no root cause could be established.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.If a sample is returned at a later date, the investigation will be reopened and the sample investigated.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7695579
MDR Text Key114228203
Report Number1644019-2018-00170
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number8065751613
Device Lot Number1950120H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-